Wednesday, 20 May 2015

Earth leakage current Test according to (IEC 60601-1)


What is leakage current :-  

High voltage is generated in an electrical appliance that use a chimerical power supply as its main power source. Should an abnormality occur in such an appliance(due to faulty design or fault), electric current flows through the body of anyone who touches the appliance. This electric current is called “ leakage current.”

Earth leakage current:-

The electric current that flow through a protective grounding wire to the ground is knows as earth leakage current.

Earth leakage current
How to perform:-

For class I equipment, earth leakage current is measured as shown in figure 1. The earth leakage test is valid for class 1 equipment with Type B, BF and CF applied part.
According to IEC 60601-1 standard Earth Leakage Current (8.7.3 d) .  The allowable earth leakage current to 5mA in normal condition and 10mA under single fault condition.


Procedure:-

1. As shown in the figure 1 , Connect the instrument to the equipment under test with the power cord.

2. Start earth leakage current measurement.



Pass criteria:-

1. The allowable earth leakage current to 5mA in normal condition and 10mA under single fault condition.




If you have any query regarding earth leakage current testing and Certification. Kindly send us your testing requirements with the product technical details to submit you our testing charges.
We are waiting for your earliest reply.You can call or write us for any clarification.Thanks & Regards,Puneet SharmaM: +91.8196980555E-mail:-(itcindiaone@gmail.com)ITC India Pvt. Ltd

1 comment:

  1. That is very interesting I love reading and I am always searching for informative information like this. I am very happy to your post about on. Great information, I would like to say your post is very informative.
    lab test near me

    ReplyDelete